Laser and X-ray Outlet Modules provide a convenient and safe power interface for portable laser and X-ray devices used in operating rooms. These modules work in conjunction with the Model XTL and XTLD Isolated Power Panels. A flush mount design includes a durable stainless steel front with a hinged door covering the laser or X-ray receptacle.

Model XTL Isolated Power Panels include a standard configuration includes a Programmable Logic Controller (PLC) system. This system protects against accidental overload and allows power to be selectively provided to the location needed. The outlet is energized by opening the receptacle cover, activating the contact switch and signaling the panel PLC to power the circuit. Once energized, the circuit is locked on until the device is disconnected and the cover is closed. The lockout feature can also limit the number of energized outlets, allowing the hospital to control the number of laser and/or X-ray procedures performed at any one time.

The number of energized outlets operated simultaneously is determined by the transformer capacity and the distance of outlet locations from the panel. The PLC protects against overloading the panel transformer. 


Model RLO-R/RXO-R Outlet Module
DRA-1-Remote

Model RLO-R (for use with laser) or RXO-R (for use with X-ray) Outlet Modules are used in conjunction with the XTL Panel. These Outlet Modules include a LIM Remote Annunciator. They can include a DRA-1 Remote Annunciator which contains basic audible and visual alarm indication along with a silence feature. The Oulet Module can also include a Remote Annunciator which contains all the same indication and functionality shown on the front of the LIM. This includes basic alarm indication as well as real-time THC, silence feature and system status display. These Outlet Modules will include a DRA-VS (for use with the Mark V LIM). The RLO-R/RXO-R Outlet Modules also include user specified NEMA receptacle, galvanized back box and a stainless steel front trim with hinged door over receptacle.
(Image shown to the left is Model RLO-R with a Model DRA-VS Remote Annuncaitor.)


Model LO/XO Outlet Module
LO/XO Outlet Module
Model LO (for use with laser) or XO (for use with X-ray) Outlet Modules are used in conjunction with the XTLD Panel. The LO/XO Outlet Module includes user specified NEMA receptacle, galvanized back box and a stainless steel front trim with hinged door over the laser or X-ray receptacle. This module is specified when the XTLD Panel is not located in the operating room.
(Image shown to the left is Model LO.)

Note: When laser receptacles are not installed in a wall but instead incorporated in an equipment boom, a separate Remote Annunciator is required. (See Remote Annunciators for options.) In such installations, it will be the customers responsibility to install a circuit switch. Contact PG LifeLink for more information.

read more