Product Line:

Laser and X-ray Isolated Power Systems

  • Description
  • Control Options
  • Components
  • XTL Accessories
  • Installation
  • Ordering
  • Specifications
  • Download Drawings
  •     Ordering

    Model # XFMR
    KVA
    Primary
    Voltage
    Secondary
    Voltage
    #
    Cont/CB
    Modifications
    XTL X -X X -X -X   (Add suffixes as required)
    Specify Laser
    or X-Ray
    1a - (110) 1a - (110) 1 H   Main Circuit Breaker
    1 - (120) 1 - (120) To J   Surface Mount
    5 2 - (208) 2 - (208) 12 RLO   (1-4) Remote Plug-in Control w/Lock-Out
    7-1/2 2a - (220) 2a - (220) *    (Specify # of rooms to be on at any one time 1-4)

    10 3 - (240) 3 - (240) R   Remote Plug-in Control

    15 4 - (277) RS   Remote Plug-in Control With Remote Meter

    25 5 - (480) *  

    Maximum number of circuits for plug control with lockout is twelve.


    Example:

    XTL 25 — 42 — 8 — RLO-3-Laser
    25 KVA 277 / 208 volt with 8 circuits, plug control with only 3 circuits on at any one time. Specify amperage on branch breakers.


    NOTE:

    All Models are single-phase. For three-phase panels, contact factory.

    NOTE:

    No branch circuits should be longer than 150 feet. Use low dielectric wire on power circuits.