Quality Components Ensure Quality Products
Nowhere is PG LIFELINK's commitment to quality, and above all safety, more evident than in the components that go into the manufacture of our Isolated Power Panels.
These components include:
The single phase, stacked core Isolation Transformer with electrostatic shield is designed for low leakage and quiet operation, single input voltage and dual secondary voltage.
Primary voltage 208, 220, 240, 277 or 480.
Secondary voltage 110, 120, 208, 220 or 240.
The self-testing, self-calibrating PG LIFELINK Mark IV Line Isolation Monitor offers continuous monitoring with non-switching circuitry, as well as low monitor hazard current of 25 microamps, a 5 milliamp alarm point and sensitivity to LIM "loss of ground". For more information, refer to the Line Isolation Monitor section.
In accordance with code requirements, we use two-pole Circuit Breaker Panels. The panel houses the primary main circuit breaker and up to 16 secondary breakers with a secondary 120 volt main.
The Branch Signal Terminal Blocks are supplied for the connection of remote signal conductors if required.
208 volt Secondary Circuit Breaker, two 2-pole 208 volt. Specify amperage required.
The Ground Bus is used for connection of all internal and branch ground conductors.
Notes
- Box will be made of galvanized steel in accordance with UL Standard 50.
- Bolt-in chassis fastens to rear of box.
- Front will be fabricated of 12 GA., type 304, stainless steel and finished with a #4 satin polish.
- Branch circuit wiring to be low leakage type XHHW.
- Annunciator wires to be 18 THHN or equal.
Optional Accessories
- Receptacles 120 volt, 20 amp duplex hospital grade.
- Receptacle 208 volt, specify amperage and NEMA configuration.
- Ground Jack 30 amp.